HDA 200 Audiometric Headphone
Closed dynamic headphones designed for extended high frequency testing.
Features
Excellent passive attenuation (based on Peltor TM Ear Defenders)
Very high quality sound reproduction
Convenient single sided cable
Padded headband and additional adjustable/removable cushions for increased comfort
Soft, replaceable circumaural ear pads
Color coded ear cups, right (red) left (blue)
Technical Data
Frequency response
<
20 to > 20,000 Hz
PTB calibrated
see table
Transducer principle
dynamic, closed
Nominal impedance
40 Ohm
Characterisitic SPL
100 dB at 1 kHz, 1 mW
Max permanent load
500 mW
Coupling
circumaural
Caliper pressure
10 N
Weight (with cable)
330 g
Cable approx. 3 m,
single-sided, open-ended
Connection
yellow
+ L
black
L
red
+ R
white
R
Standard
SPL
Passive
Maximum SPL
frequencies (dB 20
µPa) attenuation
(<10 min. @
(Hz) @ .5 Vrms
(dB)
5 V RMS)
Sennheiser Electronic Corporation One Enterprise Drive PO Box 987, Old Lyme, CT 06371
Tel: 860-434-9190 Fax: 860-434-1759 Web site: www.sennheiserusa.com
HDA200 is PTB approved. Physikalisch-Technische
Bundesanstalt, Braunschweig, Germany. PTB is equivalent to
the National Bureau of Standards.
125
112.5
14.3
132
± 3
250
113.0
15.9
132
± 3
500
112.0
22.5
132
± 3
750
111.0
-
131
± 3
1,000
108.5
28.6
129
± 3
2,000
104.0
-
124
± 3
3,000
104.0
32.0
124
± 3
4,000
104.0
45.7
124
± 3
5,000
106.5
-
127
± 3
6,000
107.5
-
125
± 3
8,000
105.5
43.8
125
± 5
9,000
105.0
-
123
± 5
10,000
102.5
-
122
± 5
11,200
102.0
-
123
± 5
12,500
103.0
-
118
± 5
14,000
98.5
-
119
± 5
16,000
100.0
-
120
± 5
· All measurements are done on a calibrated coupler
B&K 4153 (artificial ear) with the standard cone
YJ0304 above the adapter plate, type DB 0843.
· The pressure of the headband shall be 10N
± 1N.
· The RMS input voltage to the headphone is 0.5 V.
· The measurements are done with steady state sine
wave signals.
· The output impedance of the signal source shall be
<1 Ohm.
· Climatic conditions:
Temperature
T=20
º C
Humidity
H=50%rel
Atmospheric pressure
P=approx. 100kPa
All data are influenced by temperature, humidity and static pressure.
HDA 200 Frequency Response Test Conditions